OPTICS AND OPTICAL MEASUREMENT INSTRUMENTS
HIGH RESOLUTION FFP MEASUREMENT OPTICS FOR IR RANGE M-Scope type FHR
PRODUCT OVERVIEW
M-Scope type FHR is an optical system for measuring FFP in 1310~1550nm NIR spectral range with high accuracy. The combination with VGA Type InGaAs High Sensitivity NIR Detector ISA041VH enables high-accuracy FFP measurement with an angle pixel resolution of 0.1° in 1310~1550nm spectral range.
It can be applied to high precision FFP measurement and N.A. measurement of various optical devices in optical communication fields such as optical fibers, optical waveguides, optical modules, and Si photonics devices.
It is possible to build high precision FFP measurement system in combination with SYNOS' Optical Beam Analysis Module AP013.
FEATURE
- Specially designed optics for realtime observation and analysis FFP (Far Field Pattern) in IR range
- Realize high angle pixel resolution of approx. 0.1° in 1310~1550nm wavelength range
- Realize wide angle coverage of approx. ±32° × ±25.6° in 1310~1550nm wavelength range
- Long working distance design of approx. 6mm
- VGA Type InGaAs High Sensitivity NIR Detector ISA041VH is adopted as the designated detector
APPLICATION
- FFP, N.A. measurement of infrared laser devices and optical modules
- 1310nm~1550nm IR laser diodes for optical communication
- Optical fibers
- Optical waveguides
- Si photonics devices
- Various optical modules for optical communication
SUMMARY OF SPECIFICATION
MAIN SPECIFICATION OF THE OPTICS
Measurement method | Dedicated f-θ lens optics and image processing method | ||
---|---|---|---|
Measurement light flux diameter | approx. 1mmφ | ||
Measurement wavelength | 1310-1550nm wavelength range | ||
Working distance | approx. 6mm±0.8mm | ||
Optical attenuation method | By neutral density filters | ||
Camera mount | C mount | ||
Measurement angle coverage | Detector | Measurement angle coverage | Measurement wavelength |
ISA041VH | approx. ±32°(H)× ±25.6°(V) | 1310-1550nm | |
Pixel resolution | Detector | Pixel resolution | Measurement wavelength |
ISA041VH | approx. 0.1° | 1310-1550nm |
SIMPLE STRUCTURE OF HIGH RESOLUTION FFP MEASUREMENT OPTICS FOR IR RESION M-Scope type FHR
As shown in the figure on the left, the light flux, having incident angle θ from the sample, is focused at a point on the detector through f-θ lens, field lens and relay lens module. By this way, FFP image of the sample is formed and acquired by imaging detector, and analyzed directly and quickly by image processing method.
COMPONENT
STANDARD COMPONENT
- M-Scope type FHR (optics unit) : 1set
- Base plate for optics : 1set
OPTION
- DETECTOR
- VGA Type InGaAs High Sensitivity NIR Detector ISA041VH
- Spectral sensitivity : 950~1700nm
- Neutral density filter
- IR ND Filter Set NDF IR-5
- About neutral density filters, please refer to here.
- Optics workbench for sample measurement
- Optics Bench For Fiber Measurement OP002-F3/OP002-F5
- Vertical Type Optics Bench OP002
- About optics workbench, please refer to here.